MIRA - replacing auto focus pin

Modified on Tue, 3 Jan, 2023 at 1:26 PM

1: Loosen the black set screw on the right side of the auto-focus housing (circled in blue). 

2: Spin the gold pin counter-clockwise until the pin comes out. (The red "plunger" portion at the top of the laser head that goes through the black auto focus sensor "shoe" may fall - as it is not threaded into the auto focus pin any longer.

3: Put in a new pin and screw it into the red "plunger" mentioned above. 
4: Tighten the set screw again and test the auto focus pin now actuates up and down by following this document:
https://cancam.freshdesk.com/en/support/solutions/articles/62000222430-mira-testing-the-auto-focus-pin-sensor

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article